		<words audio="tts" file=""><![CDATA[Gently mix the sample to make sure that it is homogeneous.]]></words>
		<words audio="tts" file=""><![CDATA[Move the mixing wire to the end opposite to that from which the sample is to be
aspirated.
Note: If petroleum jelly, such as Vaseline, is used at the puncture area, introduce
the capillary sample into the analyzer from the end without petroleum jelly.]]></words>
		<words audio="tts" file=""><![CDATA[Remove the end caps of the capillary tube.]]></words>
		<words audio="tts" file=""><![CDATA[Put the Clot Catcher on the end opposite to that with the mixing wire.]]></words>
		<words audio="tts" file=""><![CDATA[Hold the capillary tube and tap the Capillary button.
The analyzer opens the inlet.]]></words>
		<words audio="tts" file=""><![CDATA[If measurement mode can be selected, select measurement mode.
Note: If you selected the wrong mode, tap the Reselect button and select the correct mode.
Note: If the Other modes button is available, tap it to get access to more modes.]]></words>
		<words audio="tts" file=""><![CDATA[Follow the instructions on the screen.]]></words>
		<words audio="tts" file=""><![CDATA[Place and hold the end with the Clot Catcher in the center of the Inlet Gasket.
Note: If you turn the capillary tube slightly when you place it in the center, it may be easier to put it in the right place.
]]></words>
		<words audio="tts" file=""><![CDATA[Carefully push the capillary tube into the analyzer as far as it will go and hold it
there.]]></words>
		<words audio="tts" file=""><![CDATA[Hold the capillary tube in the pushed-in position until the analyzer tells you to remove it.]]></words>
		<words audio="tts" file=""><![CDATA[When the analyzer tells you to, remove the capillary tube.
The analyzer closes the inlet.]]></words>
		<words audio="tts" file=""><![CDATA[If necessary, select a different report layout as follows:
a) Tap the current Report layout shown on the screen.
b) Select a new layout from the list.
c) Tap the Select button.]]></words>
		<words audio="tts" file=""><![CDATA[Enter the necessary data in the Patient identification screen.
Note: It is mandatory to enter data in fields with this icon:]]></words>
		<words audio="tts" file=""><![CDATA[If the Patient result screen opens before you have entered the necessary data, tap the ID button.]]></words>