		<words audio="tts" file=""><![CDATA[Press Stat Mode in the function toolbar.]]></words>
		<words audio="tts" file=""><![CDATA[Enter the Sample ID and press another field so that the software checks if the SID is already known.]]></words>
		<words audio="tts" file=""><![CDATA[If your instrument is connected to a Host (LIS or Yumizen P8000), press Query to retrieve the analysis data for the sample and acknowledge the tube presence.]]></words>
		<words audio="tts" file=""><![CDATA[If your instrument is not connected to a Host, enter all the information related to the blood sample and the patient.]]></words>
		<words audio="tts" file=""><![CDATA[If you have different orders associated with a same SID, the Run analysis window displays.]]></words>
		<words audio="tts" file=""><![CDATA[Select the analysis type.]]></words>
		<words audio="tts" file=""><![CDATA[If you have diluted the sample before the analysis, enter the dilution factor (from 1 to 10) so that the factor is taken into account in the results rendered.]]></words>
		<words audio="tts" file=""><![CDATA[Mixing specimen warning iconBe aware that if you have diluted the sample before the analysis, some alarms can be falsely not triggered only due to the pre-dilution.]]></words>
		<words audio="tts" file=""><![CDATA[Press Validate in the contextual toolbar.]]></words>
		<words audio="tts" file=""><![CDATA[Gently and thoroughly mix the sample.]]></words>
		<words audio="tts" file=""><![CDATA[Put the tube in the tube holder and close the door to start sampling.]]></words>
		<words audio="tts" file=""><![CDATA[Tube placement iconWhen placing your tube in the tube holder, make sure that you are using the appropriate position and that the cap can be pierced. If not, remove the cap before closing the door.]]></words>
		<words audio="tts" file=""><![CDATA[When the door opens, remove the tube from the tube holder and put the cap back on if needed.]]></words>
		<words audio="tts" file=""><![CDATA[imageIn case of rerun, it is recommended to remove the cap before closing the door.]]></words>