		<words audio="tts" file=""><![CDATA[Access: Home > Quality Assurance > Quality control]]></words>
		<words audio="tts" file=""><![CDATA[Make sure you have created the control lot before running it so that it is not run like a blood sample.]]></words>
		<words audio="tts" file=""><![CDATA[Prepare your control blood according to the specific instructions detailed in the control blood package insert.]]></words>
		<words audio="tts" file=""><![CDATA[Press Stat Mode in the function toolbar.]]></words>
		<words audio="tts" file=""><![CDATA[Enter the sample ID of the control blood sample.]]></words>
		<words audio="tts" file=""><![CDATA[Press Validate in the contextual toolbar.]]></words>
		<words audio="tts" file=""><![CDATA[Gently and thoroughly mix the sample.]]></words>
		<words audio="tts" file=""><![CDATA[Put the tube in the tube holder and close the door to start sampling.]]></words>
		<words audio="tts" file=""><![CDATA[Tube holder instruction iconWhen placing your tube in the tube holder, make sure that you are using the appropriate position and that the cap can be pierced. If not, remove the cap before closing the door.]]></words>
		<words audio="tts" file=""><![CDATA[When the door opens, remove the tube from the tube holder and put the cap back on if needed.]]></words>
		<words audio="tts" file=""><![CDATA[Mixing specimen warning iconRisk of erroneous results if the specimen is not continuously mixed between each analysis. Keep on mixing the specimen between each analysis.]]></words>
		<words audio="tts" file=""><![CDATA[To Check Control Results]]></words>
		<words audio="tts" file=""><![CDATA[Select a control lot.]]></words>
		<words audio="tts" file=""><![CDATA[Make sure that the results are within control target values range.]]></words>
		<words audio="tts" file=""><![CDATA[If results are out of range, perform a concentrated cleaning and rerun the control blood.]]></words>
		<words audio="tts" file=""><![CDATA[If results are still out of range, please check your reagents and your control blood stability and then contact your local HORIBA Medical representative.]]></words>